G a u ge B l oc k C o m pa r ato r s EN ISO 3650
Minimum profile
TESA UP – Software Programme for Value Processing requirements for the
computer needed
TESA UP programme for processing measured values suitable for both TESA to run the TESA UP
gauge block comparators UPD and UPC as well as for comparators from other software programme
manufacturers. Personal Computer
–– Choice of 10 languages. • Configuration
–– On-line processing of length and temperature values as transferred. without heat source
–– Measurement cycles and result outputs according to EN ISO 3650. to avoid disturbing
–– Flexible architecture for optimum adaptation to specific user’s needs. the ambient
–– Possible entry of limit values and accuracy grades peculiar to users. temperature at the
–– Surveillance of value dispersion or value drift throughout length and tempera- measurement spot
• Operating system:
ture measurements. Windows 7 or
–– Automatic execution of all relevant corrections. The programme makes earlier versions
(32 bits)
allowances for actual sizes of the reference standards, flattening due to • Processor: 650 MHz
different materials used (steel, tungsten carbide, ceramic), compensation of • 1 Hard disc (6 GB)
temperature variations with reference to 20°C according to the varying coeffi- • RAM capacity:
cients of linear expansion – as typical examples. 64 MB
–– Assignment of gauge blocks to their relevant grade. • CD-ROM drive (24x)
–– Possible storage of gauge block set related data. • RS232 serial port
–– Inch or metric value processing. 1 for length values
–– Calibration certificate in several formats. 1 for temperature
values
• 3 USB ports
05960025 TESA UP software programme 1 CD-ROM
for gauge block calibration plus 1 USB key of protection
Gauge Blocks for the Calibration of Comparators EN ISO 3650
To calibrate both TESA gauge block comparators UPD and UPC, we recommend the Special high-alloy
use of the gauge block set described hereafter. The 9-piece set is alsoy required steel, wear resistant
for calibrating TESA UPD. and stable.Excep-
Set composition including 11 steel gauge blocks, class K tion: 6 mm special
Each pair is in full compliance with: carbide gauge
–– EAL-G21 – Calibration of gauge block comparators – European cooperation for blocks.
The given expanded
Accreditation of Laboratories uncertainty
–– DKD-R 4-1 – Guidelines of the German Calibration Service (DKD) for the k = 3 refers to the
difference of central
calibration of gauge block comparators. length of both gauge
blocks A and B
S59110152 Set of 11 gauge blocks with PTB (Physikalisch Technische Budesanstalt) μm forming the pairs
certificate ± 0,015 1 to 5 as well as to
the deviations fo and
S59110489 Set of 11 gauge blocks with DAkkS certificate ± 0,030 fu from the central
Full tungsten carbide set also available on request length of gauge
blocks forming both
pairs 2 and 3. No
need to calibrate
those of pair No. 6.
Pairs Nominal length B mm
N° A mm 0,5
9 10 1 0,5
10
10
2 1,0 1,005
0,013±0.003 3 1,0 1,01
4 4,5 4,5
5 100,0 100,0
6 6,0 6,0 *
* Special bridge-shaped gauge blocks (see drawing) used for establishing the
measuring deviations of lower probe B.
L-8