Input impedance                   E le c t r o n i c L e n gth M easu r i n g E q uipme n t
970 ± 50Ω (13 kHz)
or 2150 ± 50Ω                   Calibration Standards – Dummy Probes
(standard 0 μm)
Phase (13 kHz):                 Calibration standards – also known as "dummy probes" – are resistance dividers.
71 ± 2°. Input                  Each calibration standard simulates a given length dimension with high accuracy. 
resistance:                     Each calibration standard has 2 values (positive and negative). The values indi-
100 ± 5Ω. Output                cated below are the nominal values. 
impedance at                    These products are calibrated and supplied with an inspection report that shows
13 kHz: 1000 ± 2Ω.              the values (actual values) measured during calibration and the related measuring
Phase (13 kHz): 0,2°            uncertainty.
Dummy probe (half-
bridge), sensivity              The calibration standards are connected to the instrument in place of regular
73,75 mV/V/m.                   probes. For the calibration and all required setting operations of the instrument,
Suitable for                    certain criteria and conditions need to be respected.Consult the user manual or
instruments with                get in touch with our specialists for further information.
following features:
Frequency: 13                       Set of 3 calibration standards (S41077249)  Value of the calibration stan-
± 0,65 kHz, Voltage:                                                            dard (microns)
3 ± 0,015 Veff (2               Dummy probe
symetrical voltages             Dummy probe                                     ±0
of 1,5 Veff) Input and          Dummy probe                                     ±3
output impedance:               Dummy probe                                     ± 100
≤ 0,2Ω et 2000Ω,                Dummy probe                                     ± 190
respectively                    Dummy probe                                     ± 300
Calibration: 40 % to            Dummy probe                                     ± 500
60 %. Operating:                Dummy probe                                     ± 1000
20 % to 80 %.                   Set of 3 dummy probes                           ± 1900
Storage: 5 % to                 Set of 2 dummy probes                           ± 0 / ± 100 / ± 1000
95 %. Without                                                                   ± 190 / ± 1900
condensation.

IP40 (IEC 60529)

Inspection report

Ø 18 mm, length
118 mm

≈ 45 g

20 ± 0,5°C, stabilisa-
tion time = 8 h

± 3 ppm/°C. Ageing:
± 30 ppm/a

                     S41078077
                     S41078079
                     S41078228
                     S41078230
                     S41078087
                     S41078332
                     S41078751
                     S41078752
                     S41077249
                     S41078654