Page 300 - TesaKatalogen
P. 300
DIN 32876 U S B , D C , W i r eless P r obes
Part 1
Nickel-plated Wireless Probe ± 5 mm, Large Measuring Range
housing
Stainless steel Probes developed for devices requiring a greater freedom of movement during the
measuring bolt, measurement or for parts with large dimensions.
hardened. Viton Bidirectional and wireless communication synchronized with the TWIN-STATION
= highly resistant Receiver.
fluoroelastomer –– Resolution 0,1 μm.
Fixing shank –– Range of 8 m, depending on environment.
Ø 8 mm. Measuring –– TESA wireless communication protocol independent from WiFi or Bluetooth.
bolt guided on –– Autonomy 40 hours (rechargeable battery).
ball-bearing. –– Mounting body Ø 8 mm with possoboilty of clamping over entire length.
Distance from elec- –– Measuring bolt mounted on ball bearing.
trical zero of both –– Separate guide bearing on the mounting body in order not to negatively
stops is either ad-
justable (downward) influence the movement of the measuring bolt in the event of improper
or depending on clamping on the probe body.
the position of the –– Level of protection IP54 according to IEC 60529.
lower stop (upward). –– Wide range of measurement probes.
Interchangeable –– The TWIN-STATION (part number 05030012) manages and synchronizes up
inserts. M2,5 thread. to 8 wireless probes.
Carbide ball tip –– Interface Software TIS included in supply content of the TWIN-STATION
Ø 3 mm. Connector (part number 05030012): display of measured values, possibility to indicate
Mini-jack for tolerances, simple functions +A, -A, +A+B, +A-B, and export of values in a .csv
charger. file.
Mechanical max. Note: The sales is limited to EU countries, Switzerland, USA, Canada and China.
frequency**: 60 Hz
Power supply: 208.7
100 ÷ 240 VAC,
50 ÷ 60 Hz; 240 mAh R 1,5 ø 8h6
Rechargeable
battery: 3,7 V, min. 29.6 83 10 52.8
550 mAh or 800 mAh
Frequency band: -5 +5
2,4 GHz Range: 0
8 m, depending on
environment. GT 61 W GT 61 W
Wireless transmis-
sion, TWIN-STATION
Receiver (05030012)
± 0,2 μm/°C
20 ± 0,5°C
Protection operating
enevelope
IP54 (IEC 60529)
GT 61 W: 8 g
Inspection report
with a declaration of
conformity
03230502 GT 61 W Measuring Nominal mea- Bolt Sealing
range, mm suring force*, N retraction bellows
±5 0,9 Mechanical Viton
Optional accessory:
05030012 TWIN-STATION Interface for wireless probes
Max. bolt Maximum Repeatability, Hysteresis, Setting of lower Cable output Data sheet
travel, mm permissible μm μm stop of measuring No.
10,3 error, μm 0,24 0,5 bolt***, mm
(L in mm)
GT 61 W 0,8 + 0,8· L Fixed stops lower Without cable 03200621
-5 upper +5
* Electrical zero (N) ± 25 % deviation limit. Valid in vertical mounting position, measuring bolt lowered and in static measuring.
** For an amplitude of 10 % to the last value of the measuring range.
*** Distance from electrical zero.
O-27